Key takeaways
A shock condition is locked by four values - waveform, peak, nominal duration and velocity change. Miss one and two reports stop being comparable. Peak governs how large the force is, duration governs which frequency band is excited, velocity change governs how far the part is pushed; when these three do not line up, no amount of fixture care recovers the result.
The sentence "50g, 11 ms, half-sine" carries only three pieces of information: peak, nominal duration and waveform. Keep the peak and change the duration and it is a different test - a different frequency band is excited, the overall displacement changes, and the failure site may move from structural deformation to local resonance. Miss one parameter and two reports can only be explained separately, never compared.
1. A shock condition is locked by four values
This is where shock and vibration part company: vibration looks at energy accumulated over a long time, shock looks at the transient response over a very short one. A shock condition therefore has to state both how large the force is and how long, and in what shape, it acts.
Four values, none optional.
| Parameter | What it governs | What happens if it is left out |
|---|---|---|
| --- | --- | --- |
| Waveform (half-sine, terminal peak sawtooth, trapezoidal) | How energy is distributed in time, and how much high-frequency content the pulse carries | Same peak and duration, different waveform, different velocity change and response spectrum |
| Peak acceleration | Maximum acceleration at the table and mounting point; the stress ceiling for quasi-static parts | Peak without duration states only half the condition |
| Nominal duration | How long the pulse lasts; whether the part is nudged or struck | Without it, no way to tell quasi-static response from amplified response |
| Velocity change | The momentum change the part actually receives; sets overall displacement and global deformation | No way to verify the test ran as written, and no way to align with another report |
Velocity change is not a fifth independent parameter - it is computed from the first three. For a half-sine pulse it equals peak acceleration times nominal duration times a waveform-dependent factor, about 0.64 for half-sine. That relation is public mathematics, not a limit; its value is giving you a number to back-calculate and check in a report.
Two further items belong to the condition as well, even though they are often filed as remarks: axis and cycles per axis, and waveform tolerance, including stray pulses before and after the main pulse and any zero drift. Without them, the condition stays irreproducible even between two rooms of the same laboratory.
Where the condition values come from is covered by the base layer, GB/T 28046 and the ISO 16750 series; how values map to a mounting location is in mechanical load conditions follow location.
2. Same peak, different damage
Because peak only says how large the force is, not which frequency band it lands in.
The response of a part to shock splits into two regimes according to where its first natural frequency sits relative to the dominant content of the pulse. When the natural frequency is well above that content, the response is close to quasi-static: the whole part moves with the table, displacement tracks velocity change, and peak acceleration is a fair proxy for inertial loading. When the natural frequency falls inside the pulse band, the response is amplified: the table reads a modest peak while some internal structure sees a much higher local response, and damage lands locally rather than globally.
Duration is exactly what sets the spectral width. At one peak, a shorter duration pushes energy toward higher frequencies; a longer duration carries more low-frequency content and larger overall displacement. That is why "50g" alone means nothing - it does not say whether the force was a quick strike or a slow shove.
The division of labour between the three stresses is already laid out in GB/T 28046.3 mechanical load: sine sweep finds where the part is weak, random vibration exercises accumulated margin, shock exercises the transient limit. One point is worth adding: shock also has a spectrum, only this spectrum is described by waveform and duration rather than by a PSD curve.
3. Three things to check in a report, and none of them is the peak
Shock reports are usually read for one number: did the peak get there. That reading misses three things.
First, the measured waveform against the tolerance band. The nominal pulse is an ideal curve; the measured one always deviates, and the standard specifies a band rather than a point. What has to be checked is whether peak, duration and shape all sit inside the band - not whether the peak alone made it.
Second, how clean the pulse is before and after. Table bounce, a loose fixture, a small gap at the mounting face all leave stray pulses around the main one. They may not show at all in the peak reading, yet they add energy, often in a different band.
Third, whether the measured velocity change matches the nominal value. Integrate the measured waveform and compare with the value implied by the nominal condition. A wide gap means the duration or waveform was patched. Three common causes: insufficient low-frequency response in the measurement chain, especially at short durations; a low-pass filter set at the wrong corner, reading the peak high or low; and a second impact counted into the main pulse.
| Check point | What to look at | What a mismatch means |
|---|---|---|
| --- | --- | --- |
| Waveform against tolerance | Peak, duration and shape all inside the band | Parameters were not run as written; later conclusions do not hold |
| Before and after the pulse | Stray pulses, zero drift | Table bounce or a loose fixture added energy |
| Velocity change | Integrated measurement against nominal calculation | Duration or waveform did not really land; recheck the measurement chain |
| Axis and cycles | All three axes, enough cycles per axis | One axis instead of three leaves directional failures untested |
4. Four columns in the outline, three traps
Four items deserve their own columns:
- The four condition values: waveform, peak, nominal duration, velocity change, plus axis and cycles per axis, with the standard revision and the source of the tolerance band.
- Fixture and mounting: fixture stiffness, mounting face and tightening torque, added mass, control point and monitoring point. If the first resonance of the fixture falls inside the pulse band, the table level is rewritten - the same point holds in reading a random vibration PSD.
- Monitoring and acceptance: powered or not, what is monitored (interruptions, resets, parameter drift), and whether the decision is taken during or after application.
- Before-and-after comparison: appearance, functional self-check, contact resistance and insulation, cross-sectioning where needed.
Three traps, each more expensive than the last:
Trap one: treating the peak as the condition. The outline says "50g" and leaves duration and waveform to whoever runs it. Once the defaults differ, two reports cannot be compared, and afterwards nobody can tell whether the difference came from the sample or the condition.
Trap two: substituting vibration for shock. Random vibration exercises accumulated fatigue; shock exercises the transient limit. Driver, mechanism and time scale all differ, so neither covers the other. How to choose between sine and random is in this selection guide, and why mechanical shock is a different test from thermal shock is in that article - thermal shock exercises expansion mismatch, not inertial loading.
Trap three: a normal function check after the test as the only criterion. Interruptions, resets and latch-up happen while the pulse is applied; a power-on check afterwards looks perfectly normal. If monitoring is not written into the outline, the report says the function was normal, and that sentence covers none of the interruptions that occurred during the test. Fixture detail and monitoring options are described under vibration and mechanical load testing and materials and mechanical testing.
How to group items into a programme is covered in the environmental reliability checklist; this article only pins down the four parameters of one shock condition.
A few plain words
The comparability of a shock report rests on the set of four parameters plus the fixture. Vague conditions cannot be rescued by a careful fixture, because the reader never learns which test was actually run.
When TMC LAB takes on a shock commission, waveform, peak, nominal duration and axis count are confirmed first, then the tolerance band and monitoring items are set. Conditions and limits follow the standard text, the customer specification and the technical agreement, with no values preset before the test. Velocity change is reported from integrating the measured waveform, side by side with the nominal value, so you can back-calculate and check it.
Stating the mounting attitude, the actual fastening method and the failure mode you care about when you send samples lets the outline be aligned in one pass. Scheduling is arranged per project; you can state your expected delivery date when submitting samples, and we will confirm a feasible plan against the current schedule and report progress at each milestone.